@misc{tilstone_radiometric_field_2025, author={Tilstone, Gavin H.,Jordan, Thomas M.,Aurin, Dirk,Białek, Agnieszka,Deru, Alexis,Ramsay, Ashley,Hieronymi, Martin,Dall’Olmo, Giorgio,Ligi, Martin,Kovach, Charles,Ansko, Ilmar,Ondrusek, Michael,Vabson, Viktor,Zibordi, Giuseppe,Gossn, Juan Ignacio,Kwiatkowska, Ewa,Vendt, Riho}, title={Radiometric field inter-comparison of fiducial reference measurements using an open source community processor}, year={2025}, howpublished = {journal article}, doi = {https://doi.org/10.1364/OE.551042}, abstract = {A radiometric field intercomparison was conducted at the Acqua Alta Oceanographic Tower (AAOT) in the Adriatic Sea from 14 to 21 July 2022 to assess differences in the accuracy of above-water radiometer systems (Sea-Bird HyperSAS, pySAS, TriOS-RAMSES) processed using an open-source community processor (HyperCP). Class-based and sensor-specific characteristics of the radiometers were used to determine the quantities Ed(λ), Lsky(λ), Lt(λ) and Rrs(λ), and their associated uncertainties. Using sensor-specific characteristics, the differences among systems were 2% for Ed(λ), Lsky(λ), Lt(λ) and 2.5% for Rrs(λ) and the uncertainties were 1.5%, 2%, 1.5% and 5%, respectively. The differences were higher when using class-based characteristics. Lwn(λ) values were also compared to the above-water AERONET-OC SeaPRiSM and in-water HyperPro II. For SeaPRiSM, the differences and uncertainties were <5% over blue and green bands when using Mobley [App. Opt. 38, 7442 (1999) [CrossRef] ] sea surface reflectance factors (ρ) and no NIR correction. For HyperPro II the differences were larger but were reduced when Rrs(λ) from the above-water systems was computed using ρ from Zhang et al. [Opt. Express 25, A1 (2017) [CrossRef] ] with similarity (sim) spectrum (spec) NIR correction. HyperCP using sensor-specific characteristics is highly recommended to reduce the associated uncertainties and to produce the highest quality data for satellite OC validation.}, note = {Online available at: \url{https://doi.org/10.1364/OE.551042} (DOI). Tilstone, G.; Jordan, T.; Aurin, D.; Białek, A.; Deru, A.; Ramsay, A.; Hieronymi, M.; Dall’Olmo, G.; Ligi, M.; Kovach, C.; Ansko, I.; Ondrusek, M.; Vabson, V.; Zibordi, G.; Gossn, J.; Kwiatkowska, E.; Vendt, R.: Radiometric field inter-comparison of fiducial reference measurements using an open source community processor. Optics Express. 2025. vol. 33, no. 7, 15756-15781. DOI: 10.1364/OE.551042}}